Skip to main content

Table 3 Connecting letters report of Tukey’s HSD test of the cross effect of glove type and solvent type. Levels not connected by the same letter are significantly different

From: Medical glove durability during exposure to different solvent agents: an ex-vivo experimental study

Glove Type, Solvent

     

Least Squares Mean

Sensicare Micro (polyisoprene, surgical), PBS

A

B

   

651.0

Sensicare Micro (polyisoprene, surgical), 70% Ethanol

A

    

649.8

Sensicare Micro (polyisoprene, surgical), DI Water

A

B

C

  

610.8

Sensicare Neoprene (polychloroprene, surgical), 70% Ethanol

A

B

C

D

 

563.6

Sensicare Neoprene (polychloroprene, surgical), DI Water

A

B

C

D

E

350.2

Triumph Micro (latex, surgical), 70% Ethanol

 

B

C

D

E

295.8

Aloe Touch (latex, exam), DI Water

  

C

D

E

274.4

Aloe Touch (latex, exam), 70% Ethanol

   

D

E

259.6

Triumph Micro (latex, surgical), DI Water

    

E

194.0

Sensicare Micro (polyisoprene, surgical), Air

    

E

175.8

U.S. Medical Glove (nitrile, exam), DI Water

    

E

141.4

Sensicare Neoprene (polychloroprene, surgical), PBS

    

E

122.4

U.S. Medical Glove (nitrile, exam), 70% Ethanol

    

E

91.6

U.S. Medical Glove (nitrile, exam), Air

    

E

88.8

Aloe Touch (latex, exam), PBS

    

E

84.4

Triumph Micro (latex, surgical), PBS

    

E

74.7

Triumph Micro (latex, surgical), Air

    

E

65.7

Sensicare Neoprene (polychloroprene, surgical), Air

    

E

53.3

U.S. Medical Glove (nitrile, exam), PBS

    

E

29.4

Aloe Touch (latex, exam), Air

    

E

24.2